The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction tubing leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot f111 was conducted.There were no non-conformances.This lot met all release requirements.A review of f111 for the reported issue shows no trends.Trends were reviewed for complaint categories tubing leak, alarm #16: collect pressure.No trends were detected for each complaint category.The complaint photographs were provided for analysis.The tubing leak reported by the customer has been verified as the provided photographs showed that the leak appears to be coming from a tubing line.However, the nature of the tubing damage cannot be determined through inspection of the photographs.Cellex kits are 100% leak tested prior to packaging and release; therefore, it is unlikely the reported tubing leak issue was present during product creation.No manufacturing related defects were identified during the evaluation.The cause of the tubing leak during the treatment could not be determined." investigation complete.(b)(4).
|