• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER MEDICAL-KALAMAZOO CUB PEDIATRIC CRIB- HYD BASE; BED, PEDIATRIC OPEN HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDICAL-KALAMAZOO CUB PEDIATRIC CRIB- HYD BASE; BED, PEDIATRIC OPEN HOSPITAL Back to Search Results
Catalog Number FL19H
Device Problems Device Slipped (1584); Naturally Worn (2988)
Patient Problem No Patient Involvement (2645)
Event Date 08/18/2017
Event Type  malfunction  
Event Description
It was reported that the siderail won't latch due to a worn siderail assembly.No patient involvement and no adverse consequence or clinically relevant delay in treatment was reported.
 
Manufacturer Narrative
It was found that the siderail would not latch/lock in the middle position and this was due to a broken/damaged latching assembly as well as a broken/damaged column.
 
Event Description
It was reported that the siderail won't latch due to a worn siderail assembly.No patient involvement and no adverse consequence or clinically relevant delay in treatment was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUB PEDIATRIC CRIB- HYD BASE
Type of Device
BED, PEDIATRIC OPEN HOSPITAL
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
chanda burghard
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key6869503
MDR Text Key86737336
Report Number0001831750-2017-00403
Device Sequence Number1
Product Code FMS
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue NumberFL19H
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/18/2017
Initial Date FDA Received09/15/2017
Supplement Dates Manufacturer Received08/18/2017
Supplement Dates FDA Received10/07/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/11/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-