Model Number 806455 |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/23/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The customer had tubing installed during calibration.They tried multiple times to close the occluder using the occlusion knob and the close button but was unable to fully close the tubing.There was a slight flow of the priming liquid.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the occluder would not fully close the tubing.Forceps were used to fully occlude the tubing.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
Evaluation is in progress, but not yet concluded.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed leakage occurred through the closed occluder head and the voltages at tp1 and tp2 were outside of specifications.The service repair technician (srt) performed calibration of pressure circuitry and verification / release test on the occluder head.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Search Alerts/Recalls
|