• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DORNOCH ULTRA DUO HIGH FLUID CART NON-STERILE INSTRUCTIONS FOR USE; APPARATUS, SUCTION, WARD USE, PORTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DORNOCH ULTRA DUO HIGH FLUID CART NON-STERILE INSTRUCTIONS FOR USE; APPARATUS, SUCTION, WARD USE, PORTABLE Back to Search Results
Model Number N/A
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/29/2017
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Pma/510k number: k081047; k123188; k133786.Once the investigation has been completed, a follow up mdr will be submitted.
 
Event Description
It was reported that the unit was leaking fluid from bottom of cart during surgery.No adverse events were reported as a result of this malfunction.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The device history record (dhr) for ultra duo flex fluid cart serial number (b)(4) was reviewed and noted no related non-conformances, requests for deviation (rfd), change notices (cn), or any other issues with manufacturing.The dhr review found no issues with the device and all verifications, inspections, and tests were successfully completed.Using crm to query for serial number (b)(4), the device was noted to have not been previously repaired by zimmer biomet surgical.The reported event was not confirmed by the service technician who performed the repair per cl ¿ repair carts rev.3.On (b)(6) 2017, it was reported from (b)(6) hospital that a cart was leaking fluid from the bottom of the unit.On 29 august 2017, rep-lite was contacted about the cart and dispatched a service technician to be at the site.The technician arrived at the site and was unable to confirm that the cart was leaking, but did find that the cart was giving drain errors and tandem errors.He replaced the level sensor in cylinder #2 with duo quad level sensor service kit (part # 91584, lot code 0025973) and the bumper mount with a car skirt mount kit (part # 76022, lot code 0022575) and then verified that the cart was functioning as intended.The technician then returned the cart to service without further incident.The device was tested, inspected, and repaired as per cl ¿ repair carts rev.4.Service work order (b)(4) on 29 august 2017.The root cause for the unit leaking fluid could not be determined.The reported event was not confirmed during inspection of the device.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRA DUO HIGH FLUID CART NON-STERILE INSTRUCTIONS FOR USE
Type of Device
APPARATUS, SUCTION, WARD USE, PORTABLE
Manufacturer (Section D)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer (Section G)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key6872364
MDR Text Key86719907
Report Number0001954182-2017-00028
Device Sequence Number1
Product Code JCX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PSEE H10
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberULDU500
Device Lot NumberN/A
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/29/2017
Initial Date FDA Received09/18/2017
Supplement Dates Manufacturer Received11/01/2017
Supplement Dates FDA Received11/01/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/07/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-