Device Problems
Detachment Of Device Component (1104); Use of Device Problem (1670)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/25/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
A ge healthcare investigation has been initiated and is ongoing.A follow-up report will be submitted once the investigation has been completed.Due to country specific privacy laws, patient information was not provided by the user.Device evaluation anticipated, but not yet begun.
|
|
Event Description
|
On (b)(6) 2017, the technologist at (b)(6) in (b)(6) reported that as they were angulating the table on their prestige si fluoroscopy system during a patient exam, the foot rest detached from the table railing and the patient fell on their back.There was no injury to the patient as a result of the foot rest detaching from the table.
|
|
Manufacturer Narrative
|
The ge field engineer (fe) visited the site for troubleshooting and found the foot rest was detached from the table railings and lying on the floor with both foot rest locks in the unlocked position.The fe inspected the foot rest and verified the locks were functioning correctly and confirmed there was no damage to the foot rest.The foot rest was installed back on the system and the system was turned over for clinical use.Ge healthcare¿s investigation has been completed and the cause of the foot rest detaching from the table railing was due to the user not installing the foot rest to the table railing correctly.The user was reminded to check and verify the locking mechanism before each use.No further actions are needed.
|
|
Search Alerts/Recalls
|