Model Number 7209807-
Device Problem Melted (1385)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/18/2017
Event Type  malfunction  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during set up, the device was so hot that it melted and came apart.It was sterilized and cleaned.No patient involved.
 
Manufacturer Narrative
Evaluation summary post market vigilance (pmv) led an evaluation of one device.Complaint of motor failure was confirmed.Cause of motor failure is a defective motor/gearbox.Motor is shorted out and damaged electronic components on the main pcb of the test control unit.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all manufacturer's quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUCLEAR
Type of Device
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
MITG - OKLAHOMA CITY
75 s. meridian ave
oklahoma city OK 73107
Manufacturer (Section G)
MITG - OKLAHOMA CITY
75 s. meridian ave
oklahoma city OK 73107
Manufacturer Contact
sharon murphy
60 middletown ave
north haven, CT 06473
2034925267
MDR Report Key6882200
MDR Text Key87510008
Report Number1643264-2017-21095
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7209807-
Device Catalogue Number7209807-
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/23/2017
Initial Date FDA Received09/21/2017
Supplement Dates Manufacturer Received01/04/2018
Supplement Dates FDA Received01/22/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-