• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. MODULAR FEMORAL STEM PRESS-FIT PLASMA SPRAYED CEMENTLESS SIZE 15; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. MODULAR FEMORAL STEM PRESS-FIT PLASMA SPRAYED CEMENTLESS SIZE 15; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Information (3190)
Event Date 08/25/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).Multiple mdr's were submitted for this event.Please see reports: 0001822565-2017- 06570 ,0001822565-2017-06571.(b)(4).Report source: (b)(6).The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Not returned to manufacturer.
 
Event Description
It was reported that foreign debris was found inside the sterile barrier.Surgery was delayed for 31-60 minutes.No further information has been made available at this time.
 
Manufacturer Narrative
Reported event was confirmed by review of the provided photo.Form the received photos, it was determined that the polybag has the black stain.Dhr was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required.The root cause can be design deficiency and transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODULAR FEMORAL STEM PRESS-FIT PLASMA SPRAYED CEMENTLESS SIZE 15
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6886127
MDR Text Key87391748
Report Number0001822565-2017-06571
Device Sequence Number1
Product Code LWJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PK063251
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00771301500
Device Lot Number63068962
Other Device ID NumberSEE H10 NARRATIVE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/25/2017
Initial Date FDA Received09/22/2017
Supplement Dates Manufacturer Received03/27/2018
03/27/2018
Supplement Dates FDA Received03/27/2018
03/27/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/27/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-