BATTLE CREEK EQUIPMENT THERMOPHORE Back to Search Results
Model Number 055
Device Problems Thermal Decomposition of Device (1071); Fire (1245)
Patient Problem No Information (3190)
Event Type  malfunction  
Event Description
Customer stated "caught fire at control switch".Product was returned for investigation.Upon further investigation into the customers complaint, the inspector found that the cord had a small burn right by the strain relief.This pad was manufactured in 2007.The product was approx.9 years old when the incident occurred.Based on the bce review of feedbacks, bce did not observe a similar issue within the lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOPHORE
Type of Device
THERMOPHORE
Manufacturer (Section D)
BATTLE CREEK EQUIPMENT
702 s reed rd
fremont IN 46737
Manufacturer Contact
uday bahal
MDR Report Key6889190
MDR Text Key87315920
Report Number1832415-2017-06752
Device Sequence Number1
Product Code IRT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 10/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator No Information
Device Model Number055
Device Lot Number072860
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/24/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-