Evaluation summary: post market vigilance (pmv) led an evaluation of one device.An evaluation was performed and showed particles under the distal cover glass, distal tip damage, fiber damage, moisture in the scope, , and cosmetic refurbishing of the scope is needed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Damage may occur if the hysteroscope is handled roughly during use.Excessive forces may result in traumatic insertion, chipping particles of the optical glass, or other damage to the unit.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|