MERGE HEALTHCARE MERGE HEMODYNAMICS; PROGRAMMABLE DIAGNOSTIC COMPUTER Back to Search Results
Model Number MERGE HEMODYNAMICS 9.40.1
Device Problem Application Network Problem (2879)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/29/2017
Event Type  malfunction  
Manufacturer Narrative
It was found during troubleshooting efforts that the customer's reported problem affected two (2) related sites because they share the same server.However, the procedures in progress prior to this issue were not affected.The hemo users could not open or create a new study, so the site decided to cancel the remaining scheduled/elective procedures and activate their stemi divert plan.Only one (1) procedure was directly affected by this issue and the procedure was completed successfully.An internal investigation is underway (hemo-11038) and it was determined that the cause of the problem occurred when a site hemo user made changes to the hemo system config that affected the procedure documentation/reporting templates component.Merge healthcare has corrected the customer's templates component so that future occurrences do not occur.To date, no other customers have been identified with this issue.Device labeling, hemo-5303 v9 user manual, addresses the potential for such an occurrence by statements such as, "caution: system config should be managed by trained support personnel or the site administrator." (b)(4).
 
Event Description
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the merge hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the merge hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the merge hemodynamics hemo monitor pc.On august 29, 2017, a customer reported to merge healthcare that a critical error occurred on the client pc during a coronary stent placement procedure.Subsequently, the hemo monitor and client pc were both rebooted causing a loss of physiological monitoring and patient data capture.Information obtained from the customer revealed that upon reboot, the client pc did not function as intended but the hemo monitor was capturing the patient's invasive pressures.The medical staff employed manual charting of the invasive pressures during the procedure while a portable monitoring device was used to capture the patient's vitals.With merge hemo not capturing physiological data, there is a potential for delay in treatment that could cause harm to the patient.However, the procedure was completed successfully through the combined use of merge hemo, a portable vitals monitor, and manual charting.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE HEMODYNAMICS
Type of Device
PROGRAMMABLE DIAGNOSTIC COMPUTER
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
meg mucha
900 walnut ridge drive
hartland, WI 53029
2623670700
MDR Report Key6899801
MDR Text Key88074018
Report Number2183926-2017-00180
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082421
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 08/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMERGE HEMODYNAMICS 9.40.1
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/29/2017
Initial Date FDA Received09/27/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/24/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-