Catalog Number 223710000
Device Problem Cut In Material (2454)
Patient Problem No Information (3190)
Event Type  Injury  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that upon post surgical examination, liner trial was found to have many deep damaging surface gouges in the instrument which could lead to sterilization problems.
 
Manufacturer Narrative
Examination of the returned instrument confirmed the complaint.Review of the provided photographs confirmed the tip damage.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRLC SCR TRL LN NEUT 28ID 50OD
Type of Device
HIP INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key6905502
MDR Text Key87807994
Report Number1818910-2017-25763
Device Sequence Number1
Product Code LXH
UDI-Device Identifier10603295106609
UDI-Public10603295106609
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 09/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number223710000
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/01/2017
Initial Date FDA Received09/29/2017
Supplement Dates Manufacturer Received10/31/2017
Supplement Dates FDA Received10/31/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-