• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MARLBOROUGH UPHOLD¿ LITE; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MARLBOROUGH UPHOLD¿ LITE; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN Back to Search Results
Model Number M0068318170
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/07/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that an uphold¿ lite was used during a sacrospinous ligament (ssl) fixation procedure performed on (b)(6) 2017.According to the complainant, during the procedure, after multiple attempts to catch the needle with the capio cage, the cage broke such that it partially detached.The carrier was also noted to be broken.The broken section of the carrier was retrieved from the patient by a hand.The procedure was completed with another uphold¿ lite.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
A visual examination of the returned uphold lite device revealed that the cage was lifted and bent.The carrier is not broken or detached.The mesh assembly was not returned for analysis.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.The investigation concluded that this complaint is associated with a product that meets the design and manufacture specifications but due to anatomical/procedural factors encountered during the procedure, performance of the device was limited.The most probable root cause classification is operational context.
 
Event Description
It was reported to boston scientific corporation that an uphold lite was used during a sacrospinous ligament (ssl) fixation procedure performed on (b)(6) 2017.According to the complainant, during the procedure, after multiple attempts to catch the needle with the capio cage, the cage broke such that it partially detached.The carrier was also noted to be broken.The broken section of the carrier was retrieved from the patient by a hand.The procedure was completed with another uphold lite.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UPHOLD¿ LITE
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key6913316
MDR Text Key88111426
Report Number3005099803-2017-02934
Device Sequence Number1
Product Code OTP
UDI-Device Identifier08714729839200
UDI-Public08714729839200
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K122459
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/17/2020
Device Model NumberM0068318170
Device Catalogue Number831-817
Device Lot Number0000056829
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/11/2017
Initial Date FDA Received10/04/2017
Supplement Dates Manufacturer Received10/12/2017
Supplement Dates FDA Received11/07/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age56 YR
Patient Weight52
-
-