|
Model Number H749236310020 |
Device Problems
Fluid/Blood Leak (1250); Hole In Material (1293)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that the outer shaft on the burr had a small hole and leak occurred.The target lesion was located in the moderately tortuous and moderately calcified proximal left anterior descending artery.A 1.25mm rotalink¿ plus was selected for use.After the procedure, when the burr was changed to a bigger one on the table with ongoing flushing, a small hole on the outer shaft of the burr was noted and was leaking.The procedure was completed with a 1.50mm rotalink¿ plus.No patient complications were reported and patient's status was stable.
|
|
Manufacturer Narrative
|
Device evaluated by manufacturer: the rotalink burr catheter was returned for evaluation as the advancer unit was disposed of at the facility.The handshake connection, sheath, coil and burr were microscopically and visually inspected.Blood was observed in the sheath.Three holes were noted on the sheath at 20cm, 20.2cm and 22.5cm from the strain relief respectively.The holes were consistent with the hemostasis valve being over tightened during the procedure.Functional testing was performed by connecting the rotablator rotalink burr catheter device up to a test advancer and then to the rotablator control console system.When the drip line was turned on, water started leaking from the holes in the sheath.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|
|
Event Description
|
It was reported that the outer shaft on the burr had a small hole and leak occurred.The target lesion was located in the moderately tortuous and moderately calcified proximal left anterior descending artery.A 1.25mm rotalink¿ plus was selected for use.After the procedure, when the burr was changed to a bigger one on the table with ongoing flushing, a small hole on the outer shaft of the burr was noted and was leaking.The procedure was completed with a 1.50mm rotalink¿ plus.No patient complications were reported and patient's status was stable.
|
|
Search Alerts/Recalls
|
|
|