• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SPECTRANETICS SPECTRANETICS LEAD LOCKING DEVICE; LLD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPECTRANETICS SPECTRANETICS LEAD LOCKING DEVICE; LLD Back to Search Results
Model Number 518-062
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Code Available (3191)
Event Date 09/13/2017
Event Type  Injury  
Event Description
It was reported that during a lead extraction procedure to remove 3 leads (ra, rv, lv), an lld device was cut and capped within a lead being extracted.Reportedly, extraction began with the rv lead using an lld, a glidelight, and a visisheath device.Significant binding was experience near the clavicle, so the surgeon switched to the ra lead.The same method was attempted on this lead, with the same result.Again the rv lead was attempted, also without success due to binding.At this point the lead was showing signs of breaking, so both the ra and rv leads were cut and capped with with the lld devices still inside.The patient was then scheduled for a surgical procedure at a later date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS LEAD LOCKING DEVICE
Type of Device
LLD
Manufacturer (Section D)
SPECTRANETICS
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS
Manufacturer Contact
nick mcnabb
9965 federal drive
colorado springs, CO 80921
719447-257
MDR Report Key6917107
MDR Text Key88422120
Report Number1721279-2017-00222
Device Sequence Number1
Product Code DRB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142116
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date12/15/2018
Device Model Number518-062
Device Catalogue Number518-062
Device Lot NumberFLP16M09A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/13/2017
Initial Date FDA Received10/05/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/09/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age58 YR
Patient Weight136
-
-