A siemens healthcare diagnostics inc.(siemens) technical support technician (tst) analyzed the event to determine the cause of the high, out of range recovery of d-dimer control level 2 on the sysmex ca-1500 system.The tst determined that the quality controls (qc) ranges were incorrectly set on the system.On (b)(6) 2017, the customer calibrated a new reagent lot and started new qc lots.The customer did not update the system's quality control file with the new lots' table of assigned values (tav) ranges.Using the previous lots' tav ranges, the customer ran quality controls and patient samples.The customer ran the patient sample and reported the result when d-dimer control level 2 recovered out of range.As per the tst's instructions, the customer replaced the lamp, recalibrated the assay, corrected the qc ranges and ran the qc, which recovered within expected ranges.The cause of the high, out of range recovery of d-dimer control level 2 is user error.The system is performing according to specifications.No further evaluation of this system is required.Mdr 9610806-2017-00110 was filed for the same event.
|