ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) FORTIFY ASSURA DR ICD, US; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD2357-40C
Device Problem Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/02/2017
Event Type  malfunction  
Event Description
It was reported that during the procedure for lead revision, the physician had difficulties unscrewing the setscrew, but was able to loosen the setscrew and reposition the lead.Upon plugging the right ventricular lead back into the device, the setscrew would not tighten.A new wrench was used and the same issue was observed.The wrench was tested on the atrial setscrew and worked appropriately.The device was explanted and replaced and the lead repositioned successfully.No symptoms were reported.The patient will continue to be monitored.
 
Manufacturer Narrative
The reported inability to tighten the set screw was confirmed in the laboratory.The septum was removed and the setscrew was found to be stripped.The cause of reported event was due to stripped stet screw.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY ASSURA DR ICD, US
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key6926052
MDR Text Key88776143
Report Number2938836-2017-32632
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2018
Device Model NumberCD2357-40C
Device Lot NumberA000020546
Other Device ID Number05414734508063
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/02/2017
Initial Date FDA Received10/07/2017
Supplement Dates Manufacturer Received05/08/2018
Supplement Dates FDA Received05/22/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/04/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
2088TC/52, (B)(4)
Patient Age44 YR
-
-