Model Number M00550601 |
Device Problem
Incorrect Measurement (1383)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/12/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.(b)(4).
|
|
Event Description
|
It was reported to boston scientific corporation that an alliance inflation syringe was used during an esophageal dilatation procedure performed on (b)(6) 2017.According to the complainant, during the procedure, the gauge needle went up until 3 atm when pressure was applied but was unable to increase above that.The procedure was completed with another alliance inflation syringe device.There were no patient complications reported as a result of this event.
|
|
Manufacturer Narrative
|
A visual examination of the complaint device revealed that the needle was at 3 atm.A functional evaluation was performed; it was noted that the gauge did not read well and returned to 3 atm after.Based on the condition of the returned device, the reported defect of gauge reading inaccurate was confirmed.The noted defect likely occurred due to handling of the device or portion of the device without direct patient contact either during shipping, unpacking or preparation of the device for the procedure.This could have led to the gauge receiving a shock causing damage to the internal mechanism of the gauge which would result in the gauge not operating to its specifications.Therefore, the most probable root cause of this complaint is handling damage.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.A search of the complaint database revealed that no other complaints exist for the specified lot.
|
|
Event Description
|
It was reported to boston scientific corporation that an alliance inflation syringe was used during an esophageal dilatation procedure performed on (b)(6) 2017.According to the complainant, during the procedure, the gauge needle went up until 3 atm when pressure was applied but was unable to increase above that.The procedure was completed with another alliance inflation syringe device.There were no patient complications reported as a result of this event.
|
|