• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. REVEAL LINQ; DETECTOR AND ALARM, ARRHYTHMIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. REVEAL LINQ; DETECTOR AND ALARM, ARRHYTHMIA Back to Search Results
Model Number LNQ11
Device Problems Migration or Expulsion of Device (1395); Communication or Transmission Problem (2896)
Patient Problems Dyspnea (1816); Discomfort (2330); Diaphoresis (2452)
Event Date 07/18/2017
Event Type  Injury  
Manufacturer Narrative
This information is based entirely on journal literature.Without a lot number or device serial number, the manufacturing date cannot be determined.Since no device id was provided, it is unknown if this event has been previously reported.Request for additional information will be made and upon receipt a supplemental report will be submitted accordingly.Referenced article: internal migration of an implantable loop recorder into the pleural space.Heart rhythm case reports (2017).
 
Event Description
A journal article was reviewed which contained information regarding a implantable loop recorder (ilr).The article reports that the patient felt discomfort at the implant side following the implant procedure.Thirty-five days later the patient developed left anterior pleuritic chest discomfort, accompanied by shortness of breath and diaphoresis.He presented to the emergency room where a chest x-ray revealed that the ilr had migrated into the left pleural space.The ilr was explanted.Further follow up did not yet yield any additional information.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVEAL LINQ
Type of Device
DETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6930781
MDR Text Key88919877
Report Number2182208-2017-01777
Device Sequence Number1
Product Code DSI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,literatur
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberLNQ11
Device Catalogue NumberLNQ11
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/08/2017
Initial Date FDA Received10/09/2017
Supplement Dates Manufacturer Received08/08/2017
Supplement Dates FDA Received02/08/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age78 YR
-
-