Follow-up #1 and final report submitted to update sections based on the results of investigation.Reported event: anspach emax 2 plus burr motor started smoking.Method & results: device evaluation and results: device evaluation was performed and the anspach emax 2 plus burr motor event was confirmed.Device history review: not performed as the anspach emax 2 plus burr motor is an oem product.Complaint history review: based on the device identification, the catsweb and trackwise complaint databases were reviewed from 2011 to present for similar reported events regarding anspach emax 2 plus burr motor started smoking failure of p/n: 110940, s/n: (b)(4).There have been no other similar events for the referenced serial number.Conclusions: the anspach emax 2 plus burr motor is an oem device.Upon receipt, the anspach emax 2 plus burr motor was bench evaluated by (b)(4) and the reported event was confirmed.Corrective action/preventive action: as the event did not involve a manufacturing related product problem indicating a non-conformity, adverse trend, or unanticipated hazard, no corrective action is required at this time.
|