The pigtail catheter was removed from packaging and placed on the sterile procedure table.When the catheter was flushed it was noted that the pigtail did not have side holes, the pigtail catheter only had one end hole.The catheter never came into contact with the patient.The labeling on the catheter packaging states the catheter should have 6 side holes.An image of a cordis catheter was received for analysis; however, per visual analysis of the received picture the body shaft tip of catheter could not be observed confirm the complaint.Later, one cath f5 inf pig145 110cm 6sh diagnostic catheter from this complaint was received for evaluation coiled inside a plastic bag.Per visual analysis of the device, neither side holes nor side holes¿ marks were observed on the catheter.No other anomalies found.A meeting was held with pet.The device was reviewed and the complaint was confirmed.The unit was thoroughly inspected and no side holes were visible on the unit.A device history record (dhr) review of lot 17693395 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿catheter (body/shaft) no side holes during prep¿ was confirmed through analysis of the returned device.The exact cause of the failure could not be conclusively determined during analysis; however, it was considered manufacturing-related.Therefore, a risk assessment has been initiated to assess this issue.
|