Model Number SC-2317-50 |
Device Problems
Fracture (1260); High impedance (1291); Kinked (1339); Component Missing (2306)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 09/19/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
Additional suspect medical device component involved in the event: model #: sc-2317-50 serial #: (b)(4) description: infinion cx 50 cm lead.
|
|
Event Description
|
A report was received that the leads of the patient were broken.It was reported that fourteen (14) out of thirty-two (32) contacts were registering high during an impedance check.It was also mentioned that patient could not feel stimulation on several programs.The patient will undergo a lead replacement procedure.
|
|
Manufacturer Narrative
|
Additional information was received that the patient underwent a lead replacement procedure.The patient was doing well postoperatively.Additional suspect medical device component involved in the event: model #: sc-2317-50 serial #: (b)(4) description: infiniontm cx 50 cm lead.
|
|
Event Description
|
A report was received that the leads of the patient were broken.It was reported that fourteen (14) out of thirty-two (32) contacts were registering high during an impedance check.It was also mentioned that patient could not feel stimulation on several programs.The patient will undergo a lead replacement procedure.
|
|
Manufacturer Narrative
|
Additional information was received that the scar tissue was removed during leads removal and it was assumed that the eyelets were in the scar tissue that was explanted.
|
|
Event Description
|
A report was received that the leads of the patient were broken.It was reported that fourteen (14) out of thirty-two (32) contacts were registering high during an impedance check.It was also mentioned that patient could not feel stimulation on several programs.The patient will undergo a lead replacement procedure.
|
|
Manufacturer Narrative
|
Sc-1132 (b)(4): device evaluation indicated that the ipg passed all tests performed.Sc-2317-50 (b)(4): device evaluation indicated that the leads complaints were confirmed.Visual and x-ray inspection confirmed cable fractures on the lead bodies at the kinked sections of the leads where the clik x anchors were positioned, which resulted in high impedances.No cables were exposed at the site.In addition, the lead bodies were cleanly cut into pieces during explant procedure.Sc-4318 (ln 18338771): device evaluation indicated that the clik anchor silicone materials from the eyelets were missing.
|
|
Event Description
|
A report was received that the leads of the patient were broken.It was reported that fourteen (14) out of thirty-two (32) contacts were registering high during an impedance check.It was also mentioned that patient could not feel stimulation on several programs.The patient will undergo a lead replacement procedure.
|
|
Search Alerts/Recalls
|