During troubleshooting efforts, merge technical support found that the windows update that the site's it department pushed out corrupted the model database.Merge technical support then overwrote the model database from a working system and then restarted sql (structured query language).It was confirmed that the client pc functioned as intended.A review of the customer's hemo case management within merge healthcare's internal database on 06oct2017 found that there have been no further call-ins by the customer regarding this issue.Device labeling, (b)(4) cardiology product security recommendations fact sheet, addresses the potential for such an occurrence with statements such as, "merge healthcare recommends that the windows firewall service be disabled on the merge hemo server, cath lab clients, and ancillary workstations.The windows firewall has caused communication problems and poor performance between the merge hemo server and connected workstations." for this reason, conclusions code 19 (human factors issue) was used.(b)(4).Methods code: actual device evaluated.Results code: interoperability problem (a problem with the mechanical, electrical, or communication interface between two or more separate devices or components.) conclusions code: human factors issue.
|
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the merge hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the merge hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the merge hemodynamics hemo monitor pc.On (b)(6) 2017, a customer reported to merge healthcare that the client pc hemo application shutdown during a procedure and would not allow the medical staff to log back in with domain or local accounts.Subsequently, the patient was moved to a known working lab onsite after active monitoring had been started.The customer further stated that the site had been running software updates prior to the shutdown.With merge hemo not capturing physiological data, there is a potential for delay in treatment that could cause harm to the patient.However, the procedure was completed successfully once the patient was moved to a functioning lab onsite.Reference complaint-(b)(4).
|