Information obtained from the customer revealed that the site has called in to merge technical support on two (2) other occasions reporting a similar issue.These events have also met the criteria as a reportable event and have been captured in initial mdr reports #2183926-2017-00186 and #2183926-2017-00187.For this occurrence, it was found during troubleshooting efforts between merge technical support and the customer that there was a loose cable connection as well as a bad block on the hemo monitor hard drive.Merge technical support walked the customer through the steps to correct the bad block issue by running a "chkdsk f/" command.The loose cable was reconnected and it was confirmed that the bad block had been removed and connection was re-established.Device labeling, hemo-6373 v10 user manual, addresses the potential for a loose cable connection in the (b)(4) section with statements such as, "question: why isn't the system showing any waveforms or numbers? answer: check for the green light on the pdm and ensure the link is up.Check to see if any cables are disconnected." two conclusions codes were used in this report: 19 (human factors issue - disconnected cable) and 13 (device difficult to operate - bad block on the hard drive).(b)(4).Methods code: actual device evaluated.Results code: interoperability program (a problem with the mechanical, electrical, or communication interface between two or more separate devices or components.) conclusions code #1: human factors issue.Conclusions code #2: device difficult to operate.
|
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the merge hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the merge hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the merge hemodynamics hemo monitor pc.On (b)(6) 2017, a customer reported to merge healthcare that the hemo monitor froze in two (2) of the site's cath labs.It was reported that this issue has occurred multiple times with sedated patients causing a delay in care while preparing third party monitoring equipment to complete the procedures.However, the exact dates were unknown but it was reported that there were no adverse outcomes as a result of the freezing issue.Date of event, was completed with this known occurrence date.With merge hemo not capturing physiological data, there is a potential for delay in treatment that could cause harm to the patient.However, the procedure was completed successfully using a third party monitoring system capturing invasive pressures and vitals.Reference complaint-(b)(4).
|