The device was not received for evaluation; therefore no physical analysis of the device can be performed.The batch number is unknown; therefore the manufacturing batch records for the complaint device cannot be reviewed.The device instructions for use state, "if resistance is felt during device placement, discontinue the procedure and determine the cause of resistance before proceeding.If the cause of resistance cannot be determined, remove the guidewire and device as a unit to prevent possible damage and/or complications." if there is any further relevant information provided, a follow-up medwatch report will be filed.
|
Device evaluation: a review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.The history records indicate this product was final inspection tested and was determined to be acceptable.As received, the specimen consists of one-1 hydro gw stf std s 260-035; returned coiled loose and double-bagged within "zip-lock" style poly pouches accompanied by the labeled mylar pouch film.After wiping the specimen with a gauze pad soaked with blood-bank saline, the specimen was subjected to visual and tactile examination.The specimen coating appears visually and tactilely consistent when examined at 1x - 18 inches, unaided, wet.The specimen presents skived damage to the polymer jacket exposing the metallic core wire at 3.09mm and 3.58 to 6.13cm from the distal end.Approximately 1.25cm of polymer jacket material and metallic core wire are missing from the distal end.The specimen also presents bend damage over the distal 23.5cm.The core wire fracture presents indications of a low cycle fatigue bending overload.Microscopically the specimen coating appears to be worn and abraded, consistent with clinical use.Except where noted, the specimen device appears visually and dimensionally correct.At this time it is not possible to assign a definitive root cause for the event as reported.As noted in the device instructions for use (dfu) warnings, "to prevent possible tissue damage, care should be taken when manipulating a device over a guidewire during the device's placement and withdrawal.If resistance is felt during device placement, discontinue the procedure and determine the cause of resistance before proceeding.If the cause of resistance cannot be determined, remove the guidewire and device as a unit to prevent possible damage and/or complications.Avoid manipulating and or withdrawing the hydrophilic guidewire back through a metal needle or cannulas.A sharp edge may scrape the coating or shear the guidewire.A catheter, introducer sheath or vessel dilator should replace the needle as soon as the guidewire has been inserted." the precautions section of the dfu states, "do not advance, withdraw, or torque the guidewire against resistance until the cause of the resistance has been determined.Excessive force against resistance may result in damage to the guidewire, interfacing device and/or duct." also noted in the dfu, "caution: inappropriate use of the guidewire may result in wire/tip breakage in the gastrointestinal tract." based on the information provided, clinical and/or procedural factors appear to have impacted on the event as reported.If additional relevant information is received, a follow up medwatch report will be submitted.
|