Model Number SC-2218-70 |
Device Problems
Bent (1059); Fracture (1260); High impedance (1291); Kinked (1339); Defective Device (2588)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 08/21/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional suspect medical device component involved in the event: model #: sc-2316-50 serial #: (b)(4) description: infinion 1x16 perc lead kit-50 cm.
|
|
Event Description
|
A report was received that the patients stimulator had a defective lead and was not working.It was also noted that the patient had high impedance and was experiencing inadequate stimulation.The patient underwent a lead replacement procedure wherein one lead appeared to be fractured right beyond the anchor.The patient was doing well postoperatively.
|
|
Manufacturer Narrative
|
Sc-2218-70 (sn:(b)(4)) device evaluation indicated that visual inspection found lead body was bent/kinked/damaged at the clik anchor site, 1 cm from the set screw mark.The fracture location was 22 cm from the distal end.X-ray inspection confirmed all cables were fractured.There were no exposed cables at the clik site fracture.The fractured cables resulted in the reported high impedances.Sc-4316 (ln:16445128) device evaluation indicated that the clik anchor revealed no anomalies.
|
|
Event Description
|
A report was received that the patients stimulator had a defective lead and was not working.It was also noted that the patient had high impedance and was experiencing inadequate stimulation.The patient underwent a lead replacement procedure wherein one lead appeared to be fractured right beyond the anchor.The patient was doing well postoperatively.
|
|
Search Alerts/Recalls
|