Catalog Number 1C8687 |
Device Problem
Disconnection (1171)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/22/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device was received and evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a clearlink transfer set separated.During the processing of stem cells for bone marrow transplant, the tubing separated from the luer activated device after they had attached a syringe to it to draw a sample.The customer was pulling a tc sample with a bd 3ml syringe when the tubing came out of the stopcock.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
The device was returned and an evaluation is complete.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Visual inspection was performed with no issues noted.Functional testing was performed and the tubing separated from the outlet port of the luer.Closer visual inspection of the tubing end showed there was no residual solvent bond present.The reported condition was verified.The cause could not be determined.A nonconformance has been opened to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
|
|