• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V2R
Device Problem Device Damaged Prior to Use (2284)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The scope was returned to olympus for evaluation.A visual inspection was performed and noted a cut with a sharp metal tab protruding through the bending section cover.Upon removal of the bending section cover, the sharp metal tab was found lifting.The scope was serviced and returned to the user facility.Based on similar reported events and investigation findings, the cause of the protruding / lifting skeletal metal tab could be attributed to the operator¿s technique.The original equipment manufacturer (oem) performed investigations related to this device issue.As a result, the oem has conducted a field corrective action including a distribution of ¿instructions for safe use¿ to mitigate the potential risk of patient injury.The ¿instructions for safe use¿ provides several warning statements in an effort to prevent equipment damage and patient injury.¿if any of the following conditions occur during an examination, immediately stop the examination and withdraw the endoscope from the patient.If the up/down angulation control lever does not move.If the angulation control mechanism is not functioning properly.Visually inspect the bending section for no metallic parts protruding from the bending section.Visually inspect the bending section for bends, twist, or other irregularities while the bending section remains straight.Visually inspect the bending section for abnormal bending shape, or other irregularities.Continued use of the endoscope under these conditions could result in patient injury, bleeding, and/or perforation.".
 
Event Description
Olympus was informed that during reprocessing, the scope was found damaged.There was no patient involvement with the reported scope.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
connie tubera
2400 ringwood avenue
san jose, CA 95131
408935-512
MDR Report Key6956987
MDR Text Key89878484
Report Number2951238-2017-00685
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK072957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date 10/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberURF-V2R
Device Catalogue NumberURF-V2R
Device Lot NumberN/A
Other Device ID Number04953170343612
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2017
Initial Date Manufacturer Received 10/03/2017
Initial Date FDA Received10/17/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Removal/Correction NumberZ-1251-2017
Patient Sequence Number1
-
-