COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number EGIATRS60AMT
Device Problems Misfire (2532); Separation Failure (2547); Failure to Form Staple (2579)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/28/2017
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter: occurred during a laparoscopic thoracic segmentectomy procedure.The stapling device was being used for resecting the tumor off of the lung.The stapler was able to fire, but the stapler did not fire all the way to the distal end.Stopped approximately four to five millimeters from the distal tip.And did not release the suture on the reinforced reloads, the material was torn loose from the reload.The staple line was incomplete at the distal end.In order to resolve the issue and complete the case, the sutures was cut with shears.There was no patient harm.The patient outcome is alive, no injury.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection and functional evaluation of the device had acceptable results.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mc dermott rd
north haven CT 06473
Manufacturer (Section G)
COVIDIEN LP LLC NORTH HAVEN
195 mc dermott rd
north haven CT 06473
Manufacturer Contact
sharon murphy
60 middletown ave
north haven, CT 06473
2034925267
MDR Report Key6961161
MDR Text Key90522102
Report Number1219930-2017-08077
Device Sequence Number1
Product Code GDW
UDI-Device Identifier20884521191010
UDI-Public20884521191010
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133938
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 01/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2020
Device Model NumberEGIATRS60AMT
Device Catalogue NumberEGIATRS60AMT
Device Lot NumberN7B0404KX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/28/2017
Initial Date FDA Received10/19/2017
Supplement Dates Manufacturer Received01/07/2018
Supplement Dates FDA Received01/25/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/13/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age64 YR
-
-