• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SELF CENT HIP 45X28 GRY; HIP OTHER IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SELF CENT HIP 45X28 GRY; HIP OTHER IMPLANT Back to Search Results
Catalog Number 103545000
Device Problem Misconnection (1399)
Patient Problems No Consequences Or Impact To Patient (2199); No Information (3190)
Event Date 09/21/2017
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that bipolar poly would not seat correctly inside of outer diameter metal head on both implants and there was movement of 2mm on the lip.
 
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
 
Manufacturer Narrative
Product complaint: (b)(4).Investigation summary: examination of the returned device is unable to determine a root cause for the problems experienced.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate. .
 
Event Description
Complaint description: it was reported that bipolar poly would not seat correctly inside of outer diameter metal head on both implants and there was movement of 2mm on the lip./ | investigation method: / | investigation summary:.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELF CENT HIP 45X28 GRY
Type of Device
HIP OTHER IMPLANT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46582
5743725905
MDR Report Key6964735
MDR Text Key90067466
Report Number1818910-2017-26997
Device Sequence Number1
Product Code KWY
UDI-Device Identifier10603295003359
UDI-Public10603295003359
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K812672
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 09/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number103545000
Device Lot Number534439
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/21/2017
Initial Date FDA Received10/20/2017
Supplement Dates Manufacturer Received10/13/2017
03/26/2018
Supplement Dates FDA Received10/24/2017
03/26/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/11/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age70 YR
-
-