BAXTER HEALTHCARE - ENGLEWOOD EM2400 DY DISPLAY; SYSTEM/DEVICE, PHARMACY COMPOUNDING Back to Search Results
Catalog Number 2400DY
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 10/03/2017
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the display cables of an (b)(4) dy display module had an exposed wire which caused connection issues.It was not specified when in the process this occurred.The customer swapped out the display with their backup, and the device was scheduled to be swapped.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The device was received for evaluation.Visual inspection revealed that the display power cable was damaged/punctured and exposing the inner shield.Functional testing was performed with no issues noted.The reported condition was verified.The cause of the condition was not determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EM2400 DY DISPLAY
Type of Device
SYSTEM/DEVICE, PHARMACY COMPOUNDING
Manufacturer (Section D)
BAXTER HEALTHCARE - ENGLEWOOD
englewood CO
Manufacturer (Section G)
BAXTER HEALTHCARE - ENGLEWOOD
14445 grasslands dr
englewood CO 80112
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key6968732
MDR Text Key90689585
Report Number1416980-2017-08491
Device Sequence Number1
Product Code NEP
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2400DY
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/03/2017
Initial Date FDA Received10/23/2017
Supplement Dates Manufacturer Received12/28/2017
Supplement Dates FDA Received01/15/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-