• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP Back to Search Results
Model Number 801041
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Patient Involvement (2645)
Event Date 09/29/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Evaluation is in progress, but not yet concluded.The reported complaint occurred when the unit was being set-up for a safety inspection.
 
Event Description
It was reported that during use of the device for a non-clinical activity, the display flickered and was unreadable.There was no patient involvement.
 
Manufacturer Narrative
(b)(4).During laboratory analysis, the product surveillance technician (pst) observed the display to flicker with missing graphics.The pst replaced the customer's graphics driver board with lab use only graphics driver board which restored functionality determining customer's graphics driver board was defective.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
katie hoyt
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key6969281
MDR Text Key90870824
Report Number1828100-2017-00486
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131618
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801041
Device Catalogue Number801041
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/04/2017
Initial Date FDA Received10/23/2017
Supplement Dates Manufacturer Received11/10/2017
Supplement Dates FDA Received12/05/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/13/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-