• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC ACE; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC ACE; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HARH36
Device Problems Loose or Intermittent Connection (1371); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The lot history records were reviewed and the manufacturing criteria were met prior to the release of this lot.
 
Event Description
It was reported that the tissue pad was loose and fell into situs during a fundoplication procedure.Part could be removed.Display showed error code reduce pressure on blade.No further information is available.
 
Manufacturer Narrative
(b)(4).Batch # p92650.Investigation summary.The analysis results found that the device was received with the tissue pad detached and not returned but with evidence of body fluids and tissue pad material in the groove section of the clamp arm.The device was connected to a test hand piece and a gen11 and the device did activate during functional testing.There were no alert screens displayed at any time during testing.The device was disassembled to inspect the internal components and no anomalies were found.Based on the condition of the tissue pad, a probable cause of this damage is that the clamp of the device may have been closed and the instrument activated without tissue present.Care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between them.Keep the clamp arm open when back cutting or while the blade is active without tissue between the blade and tissue pad to avoid damage to the tissue pad.The resulting damage contributes to the removal of the pad from the clamp arm.The cleaning of the pad, not in accordance with the ifu, can also result in removal of the pad during use.The ¿replace instrument¿ alert screen is displayed after receiving two consecutive alert screens and it is advising of a potential issue with the instrument.The batch history record was reviewed and there were no defects, protocols or ncr(s) found during the manufacturing process related to this complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC ACE
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key6970095
MDR Text Key90859252
Report Number3005075853-2017-05604
Device Sequence Number1
Product Code LFL
Combination Product (y/n)N
PMA/PMN Number
K132612
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date03/31/2022
Device Catalogue NumberHARH36
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/26/2017
Initial Date FDA Received10/23/2017
Supplement Dates Manufacturer Received09/26/2017
Supplement Dates FDA Received12/11/2017
Date Device Manufactured04/24/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-