NIHON KOHDEN CORPORATION ZM-521PA; TRANSMITTER Back to Search Results
Model Number ZM-531PA
Device Problem Failure to Analyze Signal (1539)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/07/2016
Event Type  malfunction  
Manufacturer Narrative
The biomedical engineer (bme) states that a patient had three pauses but only two alarmed.The first pause was 4 seconds long and alarmed as bradycardia, the second was 5 seconds long and alarmed asystole, the third pause was 6 seconds long and it did not alarm on the transmitter device.When reviewing in full disclosure, the customer states that in all three episodes there was a lead off but the first two still caused an alarm.Settings are all appropriate.Reviewed good electrode skin prep, placement and changing ever 24 hours.
 
Event Description
The biomedical engineer (bme) states that a patient had three pauses but only two alarmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZM-521PA
Type of Device
TRANSMITTER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn: shama mooman
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key6972937
MDR Text Key90102865
Report Number8030229-2017-00357
Device Sequence Number1
Product Code DRT
UDI-Device Identifier04931921115084
UDI-Public4931921115084
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043517
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 10/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberZM-531PA
Device Catalogue NumberZM-531PA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA10/24/2017
Distributor Facility Aware Date09/07/2016
Device Age63 MO
Event Location Hospital
Date Report to Manufacturer10/24/2017
Initial Date Manufacturer Received 10/24/2017
Initial Date FDA Received10/24/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/13/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-