• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY LUMBAR CATHETER KIT, OPEN TIP, 84 CM; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY LUMBAR CATHETER KIT, OPEN TIP, 84 CM; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Model Number 44406
Device Problems Break (1069); Device Or Device Fragments Location Unknown (2590)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 09/28/2017
Event Type  Injury  
Event Description
It was reported that while the lumbar catheter was inserted and was directed towards the spinal canal, it seem that the tuohy needle tip and the catheter tip were unable to be advanced due to being caught.Only the catheter was attempted to be removed, although the tuohy needle should have been removed as well.Then, the catheter fractured and a piece of it remained in the patient body.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received reported that the physician determined that the broken piece left inside the patient did not pose a risk to the patient and would not be likely to cause or contribute to a death or serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUMBAR CATHETER KIT, OPEN TIP, 84 CM
Type of Device
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
Manufacturer (Section G)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key6979701
MDR Text Key90324454
Report Number2021898-2017-00542
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K091312
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number44406
Device Catalogue Number44406
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/29/2017
Initial Date FDA Received10/26/2017
Supplement Dates Manufacturer Received10/26/2017
Supplement Dates FDA Received11/21/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-