• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO FOOTED ATTACHMENT 16MM; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO FOOTED ATTACHMENT 16MM; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Catalog Number 5407FA2000
Device Problems Bent (1059); Overheating of Device (1437)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 06/02/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Twenty-one events were reported for this quarter.Twenty devices were received for evaluation; eighteen events were confirmed during testing.Eighteen devices were found to be bent.The reported event was not confirmed for 2 devices; the devices were found to be within specifications for the reported event.One device was not available to stryker for evaluation.This device is not repairable and was not returned to the user facility.There were no remedial actions taken.This device is not labeled for single-use.
 
Event Description
This report summarizes 21 malfunction events in which the device had a bent foot.Fifteen events had no patient involvement; no patient impact.Five events had patient involvement; no patient impact.One event had no known patient involvement or patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOOTED ATTACHMENT 16MM
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key6981776
MDR Text Key91239291
Report Number0001811755-2017-02281
Device Sequence Number1
Product Code HBE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported21
Summary Report (Y/N)Y
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number5407FA2000
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/02/2017
Initial Date FDA Received10/26/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-