DEPUY MITEK LLC US ULTRA AGGRESSIVE PLUS 5.0 5PK;  ARTHROSCOPIC SHAVER SYSTEM BLADE, SINGLE-USE Back to Search Results
Catalog Number 283729
Device Problems Material Disintegration (1177); Metal Shedding Debris (1804)
Patient Problems Foreign Body In Patient (2687); No Known Impact Or Consequence To Patient (2692)
Event Date 09/29/2017
Event Type  malfunction  
Manufacturer Narrative
Additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
The affiliate reported via email during the surgery, the 2 devices released small particles and it was not possible to continue the surgery with them.
 
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the complaint device was received and forwarded to npd engineer for evaluation.Visual observation of the complaint device reveals no anomalies to the outer shaft.When the inner shaft was examined, the distal end showed multiple friction marks and slight damage, indicates excessive friction and indicates that the device might have hit a hard surface, or an excessive lateral force was applied to the device during use, which could lead to metal shavings as reported, confirming the complaint.A review of the device history record indicated that this batch of products was processed without incident; therefore, there is no evidence of manufacturing anomalies on the paperwork reviewed.Further, a review into the depuy synthes mitek complaints system revealed one other similar complaint for this lot of devices that were released to distribution.The only two failures identified for this lot are on this complaint.A white paper was performed previously on this failure; the result indicated that the amount of shedding for this product is acceptable.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRA AGGRESSIVE PLUS 5.0 5PK
Type of Device
 ARTHROSCOPIC SHAVER SYSTEM BLADE, SINGLE-USE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK
325 paramount drive
raynham MA 02767
Manufacturer Contact
jennifer lawrence
325 paramount drive
raynham, MA 02767
5088808100
MDR Report Key6983053
MDR Text Key90789092
Report Number1221934-2017-10633
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K041824
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 09/29/2017,01/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2022
Device Catalogue Number283729
Device Lot NumberM1704021
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2018
Is the Reporter a Health Professional? No
Event Location Hospital
Date Report to Manufacturer09/29/2017
Initial Date Manufacturer Received 09/29/2017
Initial Date FDA Received10/27/2017
Supplement Dates Manufacturer Received01/19/2018
08/27/2018
Supplement Dates FDA Received01/23/2018
08/29/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/07/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-