TERUMO CARDIOVASCULAR SYSTEMS CORPORATION LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP
|
Back to Search Results |
|
Model Number 801041 |
Device Problems
Device Stops Intermittently (1599); Device Displays Incorrect Message (2591)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/05/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Per the service repair technician (srt), there was no tubing or obstruction in the roller pump.The optical switch was replaced.The unit operates to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Event Description
|
The service repair technician (srt) reported that during routine testing of the device at the service center, the pump repeatedly stopped and displayed a 'pump jam error' message multiple times.There was no patient involvement.
|
|