MAUDE Adverse Event Report: INTUITIVE SURGICAL,INC. DA VINCI SI SURGICAL SYSTEM; ENDOSCOPIC INSTRUMENT CONTROL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL,INC. DA VINCI SI SURGICAL SYSTEM; ENDOSCOPIC INSTRUMENT CONTROL SYSTEM Back to Search Results
Model Number IS3000 A60P9B
Device Problems Electrical /Electronic Property Problem (1198); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/02/2017
Event Type  malfunction  
Manufacturer Narrative
Isi received the units involved with the complaint and completed the device evaluation.Failure analysis was unable to reproduce the reported issue.The unit was installed onto a test system and worked fine without any problems or errors.This complaint is being reported due to a da vinci system malfunction rendering the da vinci system unavailable for use after the start of a surgical procedure.Although no patient harm occurred, if this malfunction were to recur it could cause or contribute to an adverse event.
 
Event Description
It was reported that during a da vinci-assisted surgical procedure; the customer experienced a repeated recoverable fault displaying error code 45311, a 3d external monitor image issue and noise on the surgeon side console (ssc) right eye.A technical support engineer (tse) provided troubleshooting but the issue persisted.The surgeon made the decision to abort the procedure post anesthesia and port placement.No patient harm, adverse outcome or injury was reported.A field service engineer (fse) was dispatched to the facility but was unable to reproduce the reported failure.The personality module surgeon console (pmsc) was replaced to resolve the issue.The pmsc processes audio and video from the surgeon console.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DA VINCI SI SURGICAL SYSTEM
Type of Device
ENDOSCOPIC INSTRUMENT CONTROL SYSTEM
Manufacturer (Section D)
INTUITIVE SURGICAL,INC.
sunnyvale CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC.,
950 kifer rd
sunnyvale CA
Manufacturer Contact
tabitha reed
950 kifer rd
sunnyvale, CA 
4085232420
MDR Report Key6985967
MDR Text Key90763828
Report Number2955842-2017-00719
Device Sequence Number1
Product Code NAY
Combination Product (y/n)N
PMA/PMN Number
K081137
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 10/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIS3000 A60P9B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/02/2017
Initial Date FDA Received10/27/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-