|
Model Number H7493932430 |
Device Problem
Peeled/Delaminated (1454)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/27/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device evaluated by mfr: returned product consisted of a fractionated flow reserve (ffr) comet pressure wire.The tip, device shaft, sensor port and the coefficient values was examined for damage or any irregularities.The shaft showed a kink.The kink was approximately 89cm from the tip.There was also some slight peeling of the coating on the kinked area.The tip showed no damage.The sensor port was clear of any material.Inspection of the remainder of the device revealed no other damage or irregularities.The pressure wire was connected to the analysis support test bench and all applicable data was correct as designed, there was no difficulty in connecting the wire.The coefficient was confirmed to be in specification.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
|
|
Event Description
|
Reportable based on device analysis completed on (b)(6) 2017.It was reported that the comet wire was not detected.During preparation, a comet pressure guidewire was selected to perform fractionated flow reserve (ffr) measurement, however, it could not be detected by the polaris imaging system.The device was removed and the procedure was completed with a different device.No patient complications were reported.However, the returned device revealed teflon coating peeled off.
|
|
Search Alerts/Recalls
|
|
|