Siemens healthcare diagnostics has confirmed that the bc thrombin reagent kit lot 46751 (contains thrombin reagent lot 517468) and kit lot 47184 (contains thrombin reagent lot 517469) produce unexpected prolonged thrombin time (tt) results for expected normal samples, and may recover above the upper limit of normal (< 21 seconds) as stated within the ifu (instruction for use).The investigation by siemens confirmed that the reference range is shifted to an approximately 20% prolonged tt.If the reference range is not adjusted for the current affected lot, this may lead to a higher number of samples requiring follow-up in the case of slightly prolonged tt.There is a potential for misinterpreting unexpected prolonged results as unfractionated (uf) heparin contamination and some other thrombin inhibitors.An urgent field safety notice (ufsn) ph-17-018.A.Ous was sent to outside the us (ous) customers and an urgent medical device correction (umdc) ph17-018.A.Us was sent to us customers in october 2017.The ufsn and umdc remind customers that reference intervals vary from laboratory to laboratory depending on the population, the technique and reagent lot.Therefore, each laboratory must establish its own reference intervals or verify them whenever one or more of the aforementioned variables are changed.The ufsn and udmc also informs customers using the bc thrombin reagent as a screening assay for thrombin inhibition that the affected lots are more sensitive with approximately 20% prolonged/elevated tt results.This will lead to an increased rate of results above the reference range if not adjusted.Thus, the ufsn and udmc advise customers to adjust their specific reference ranges as per the application sheet for thrombin time with bc thrombin when changing to a new lot.Mdr 9610806-2017-00129 was filed for the same event.
|