Catalog Number 1805034300
Device Problem Device Slipped (1584)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/05/2017
Event Type  malfunction  
Manufacturer Narrative
The investigation identified that the issue was due to tucking the sheets under the bottom cover and not attributed to the mattress itself.The issue was resolved for the customer by discussing proper use of sheets on the mattress and explaining the importance of contact between the bottom of the mattress and the litter of the stretcher.
 
Event Description
It was alleged that the mattress was sliding off of the stretcher.There were no adverse consequences reported as a result of the mattress coming off of the litter.
 
Manufacturer Narrative
Supplemental submitted to include udi.
 
Event Description
It was alleged that the mattress was sliding off of the stretcher.There were no adverse consequences reported as a result of the mattress coming off of the litter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMFGEL SE W/O FIRE BAR, 30IN
Type of Device
MATTRESS, FLOTATION THERAPY, NON-POWERED
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
kristen canter
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key6996647
MDR Text Key92051975
Report Number0001831750-2017-00498
Device Sequence Number1
Product Code IKY
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number1805034300
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/05/2017
Initial Date FDA Received11/02/2017
Supplement Dates Manufacturer Received10/05/2017
Supplement Dates FDA Received03/23/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-