Integra has completed their internal investigation on october 25, 2017.Results: evaluation of returned device; the forceps are broken at section change at hinge.The fragment were returned.A corroded area can be observed within the fracture zone.It suggests the presence of a crack prior to the breakage.A thorough observation of the breakage area did not reveal manufacturing or material defect.Dhr review; no non-conformity related to this issue for this lot.Complaints history; no adverse trend.Conclusion: considering that no material or manufacturing defect was found and that there are evidences of a crack prior to the breakage, the most probable cause of this rupture is the recurrence of impacts on the instrument during its use, which weakened the forceps arm and progressively led to the reported event.Moreover, the ifu provided with the device requires to: "check the condition of instruments before and after each case.Remove from use any incomplete or poorly operation instruments".
|