Model Number 57063
Device Problem Crack (1135)
Patient Problem No Information (3190)
Event Date 10/12/2017
Event Type  malfunction  
Manufacturer Narrative
The customer was sent a replacement power supply.Attempts to follow up with the customer to get additional information were unsuccessful.This issue with a damaged rev m power supply for the pump in style device was addressed in investigation (b)(4).The investigation found that they were being damaged during shipment from the manufacturer to medela.This damage was causing the plastic housing to fail prematurely when subjected to normal use and foreseeable misuse.The packaging used by the manufacturer to ship the power supply to medela was not robust enough to handle all of the potential shipping, handling, and abuse conditions that could arise from logistics of the consolidation process.As a result of the investigation, the shipping and consolidation process was modified to reduce the handling and potential for double stacking of the skids.The shipping packaging strength was also increased to further protect the power supply during shipping.Since then, as a part of routine continuous improvement activities, a rev p power supply has been distributed to market, manufactured under a revised design and by a different manufacturer.
 
Event Description
On (b)(6) /2017, the customer alleged to medela (b)(4) that the power supply housing for her pump in style advanced breast pump was cracked.She indicated that it happened when they moved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP IN STYLE ADVANCED - ON-THE-GO TOTE
Type of Device
PUMP, BREAST, POWERED
Manufacturer (Section D)
MEDELA LLC
1101 corporate dr
mchenry IL 60050
Manufacturer (Section G)
MEDELA LLC
1101 corporate drive
mchenry IL 60050
Manufacturer Contact
robert sokolowski
1101 corporate drive
mchenry, IL 60050
MDR Report Key7002749
MDR Text Key92212540
Report Number1419937-2017-00308
Device Sequence Number1
Product Code HGX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
KK03614
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 11/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number57063
Device Catalogue Number57063
Device Lot Number9207010 REV M
Was Device Available for Evaluation? No
Distributor Facility Aware Date10/12/2017
Initial Date Manufacturer Received 10/12/2017
Initial Date FDA Received11/06/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N