• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 24 FR.; RESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 24 FR.; RESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS Back to Search Results
Model Number A22041A
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem Foreign Body In Patient (2687)
Event Date 10/23/2017
Event Type  Injury  
Manufacturer Narrative
The suspect medical device was not returned to the manufacturer for evaluation/investigation but to olympus medical systems corporation (omsc), (b)(4) (returned to omsc on 2017-10-23).The evaluation/investigation confirmed that the ceramic insulation at the distal end of the resection sheath had broken off and cracked.However, no part was missing since all fragments were returned as well.The cause of this damage and the breakage of the ceramic insulation is most likely thermal (excessive high-frequency output power or laser-induced) and mechanical overload in combination with wear.Furthermore, a material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the resection sheath without showing any abnormalities.It is clearly stated as a warning note in the instructions that impact, fall, shock or similar stress can damage the ceramic insulation at the sheath's distal end and that the instrument must not be used if damaged since otherwise this can cause injuries to the patient and/or user.The user apparently did not follow these instructions since the damage and breakage of the ceramic insulation were caused by thermal and mechanical overload.Therefore, this event/incident was attributed to use error and the case will be closed from olympus side with no further actions.However, the event/incident will be recorded for trending and surveillance purposes.In addition, the user will be informed about the investigation results and retrained to correctly use the olympus medical devices.
 
Event Description
Olympus was informed that during cleaning after a therapeutic transurethral resection (tur) procedure, it was noticed that the ceramic insulation at the distal end of the resection sheath had broken off and was missing.An unspecified examination was performed where fragments were noted inside the patient.These fragments were retrieved during a follow-up procedure the next day.It was confirmed that no foreign objects remained inside the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESECTION SHEATH, 24 FR.
Type of Device
RESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM   22045
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 22045
GM   22045
4940669662
MDR Report Key7005740
MDR Text Key91206313
Report Number9610773-2017-00124
Device Sequence Number1
Product Code FJL
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PK931995
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial
Report Date 11/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22041A
Device Catalogue NumberA22041A
Device Lot Number084W
Other Device ID Number04042761020961
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/23/2017
Initial Date FDA Received11/07/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/09/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
OUTER SHEATH, 26 FR., 2 STOPCOCKS (A22026A)
Patient Outcome(s) Required Intervention;
-
-