MAUDE Adverse Event Report: ENCORE MEDICAL L.P. FOUNDATION SHOULDER; GLENOID, UNIFORM, PEGGED 50MM CM FOUNDATION SHOULDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. FOUNDATION SHOULDER; GLENOID, UNIFORM, PEGGED 50MM CM FOUNDATION SHOULDER Back to Search Results
Model Number 500-04-150
Device Problem Failure To Adhere Or Bond (1031)
Patient Problem Pain (1994)
Event Date 10/09/2017
Event Type  Injury  
Event Description
Revision surgery - due to the patient being in pain from loose glenoid component.
 
Manufacturer Narrative
The reason for this revision surgery was the patient was in pain.The in-vivo length of patient service for the implant was 12 years.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.This event is deemed to be non-product related.The root cause of this complaint was a revision surgery due to the pain because of loose glenoid.There are many factors that may contribute to the event that are outside the control of djo surgical are loose joints from inadequate soft tissue support, degenerative bone, patient bone deterioration, patient activities or trauma.No additional information was submitted with the complaint regarding pre-existing conditions of the patient or any activities that may have contributed to the event and hence a definitive root cause cannot be determined.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION SHOULDER
Type of Device
GLENOID, UNIFORM, PEGGED 50MM CM FOUNDATION SHOULDER
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
Manufacturer Contact
teffany hutto
9800 metric blvd.
austin, TX 78758-5445
MDR Report Key7005928
MDR Text Key91216564
Report Number1644408-2017-00977
Device Sequence Number1
Product Code KWS
UDI-Device Identifier00888912143226
UDI-Public(01)00888912143226
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K960906
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/30/2010
Device Model Number500-04-150
Device Catalogue Number500-04-150
Device Lot Number260192
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/09/2017
Initial Date FDA Received11/07/2017
Supplement Dates Manufacturer Received11/21/2017
Supplement Dates FDA Received12/08/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/06/2005
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age85 YR
-
-