• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL,INC. DA VINCI SI SURGICAL SYSTEM; ENDOSCOPIC INSTRUMENT CONTROL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL,INC. DA VINCI SI SURGICAL SYSTEM; ENDOSCOPIC INSTRUMENT CONTROL SYSTEM Back to Search Results
Model Number IS3000 A60P9B
Device Problems Electrical /Electronic Property Problem (1198); Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/16/2017
Event Type  malfunction  
Manufacturer Narrative
Intuitive surgical inc.(isi) received the unit involved with this complaint and completed the device evaluation.Failure analysis confirmed the customer reported failure.The unit was installed and tested in the printed circuit assembly (pca) test system.Upon start-up, the unit failed with error 25581.It was determined that the issue was caused by the rdm 2 (rac drive module) was out of range and the fan did not spin.The rdm 2 and fan will be replaced to resolve the issue.This complaint is being reported due to a da vinci system malfunction rendering the da vinci system unavailable for use after the start of a surgical procedure.Although no patient harm occurred, if this malfunction were to recur it could cause or contribute to an adverse event.
 
Event Description
It was reported that during a da vinci-assisted hysterectomy procedure, the customer experienced a non-recoverable error.The customer attempted several power cycles, prior to calling intuitive surgical, inc.(isi) technical support for assistance.All arm leds turned red upon power up and the customer was unable to recover from the reoccurring fault.The isi technical support engineer (tse) noted several errors via the system logs.At that time, the surgeon made the decision to convert and complete the procedure using traditional laparoscopic techniques.There was no report of patient harm, adverse outcome or injury.An isi field service engineer (fse) was dispatched to the facility and was able to reproduce the reported failure.To resolve the issue, the fse replaced the remote arm controller board.The rac refers to the printed circuit board that receives signals from the rac control module (rcm) which performs all of the control, monitoring, communications, and upper-level safety functions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DA VINCI SI SURGICAL SYSTEM
Type of Device
ENDOSCOPIC INSTRUMENT CONTROL SYSTEM
Manufacturer (Section D)
INTUITIVE SURGICAL,INC.
sunnyvale CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC.,
950 kifer rd
sunnyvale CA
Manufacturer Contact
tabitha reed
950 kifer rd
sunnyvale, CA 
4085232420
MDR Report Key7007296
MDR Text Key92215145
Report Number2955842-2017-00755
Device Sequence Number1
Product Code NAY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081137
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Nurse
Type of Report Initial
Report Date 10/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIS3000 A60P9B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/16/2017
Initial Date FDA Received11/07/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-