• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 CONTROL PANEL MRP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 CONTROL PANEL MRP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 28-95-85
Device Problems False Alarm (1013); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/11/2017
Event Type  malfunction  
Manufacturer Narrative
Livanova (b)(4) manufactures the s5 control panel mrp.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
 
Event Description
Livanova (b)(4) received a report that an s5 control panel mrp displayed an error and an alarm sounded during a procedure.There was no tubing in the pump and the device was not in operation.The customer reset the occlusion and the device began to function as expected.There was no report of patient injury.
 
Manufacturer Narrative
No investigation based on the readout of the pump could be performed due to the age of the device and its components: this pump had the old version of the processor board and with this version it is only possible to store one error per day.As the issue could not be identified in the readout of the system, a root cause was not identified.A review of the dhr did not identify any deviations or nonconformities relevant to the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 CONTROL PANEL MRP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key7010917
MDR Text Key92517802
Report Number9611109-2017-00904
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 12/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number28-95-85
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/13/2017
Initial Date FDA Received11/08/2017
Supplement Dates Manufacturer Received11/29/2017
Supplement Dates FDA Received12/18/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/24/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-