Model Number 388928 |
Device Problems
Bent (1059); Break (1069)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/07/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information received from a healthcare professional (hcp) via a manufacturer representative reported a problem occurred during implant procedure of the lead.The mandrel was positioned while the lead was bent at the end.It was visually noticed that the mandrel had passed between the pads and changed the electrode as a precaution.It was noted that the ¿chuck¿ of the electrode passed between contacts 0 and 1 during the procedure.There was no consequence and the lead was replaced by precaution.The issue was noted as resolved.
|
|
Manufacturer Narrative
|
Analysis of the lead, model 3889-28, found lead distal end electrode distal end bent.Analysis identified the electrode at the distal end of the lead was bent.Electrical testing of the lead determined continuity was complete and no electrical shorts were identified between the circuits.Analysis identified the outer insulation of the lead was broken//torn under electrode #1.Analysis observed the distal end of the lead was stretched.Analysis observed the distal end of the lead was bent.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received reported that during the intervention, the surgeon introduced the electrode into the white dilator, then he mobilized the curved chuck in the electrode, the curved chuck then passed between contacts 0 and 1 of the electrode.
|
|
Search Alerts/Recalls
|