• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM Back to Search Results
Model Number 917000000
Device Problem Component Falling (1105)
Patient Problem No Patient Involvement (2645)
Event Date 10/19/2017
Event Type  malfunction  
Manufacturer Narrative
Additional method code: (b)(4).Maintenance review investigation: a service call was placed and a terumo bct service technician visually inspected the machine at the customer's site.The service technician was able to duplicate the reported condition.Upon visual inspection, the iv pole slowly descends upon vibration.The iv pole release button was adjusted and the machine was returned to service.An internal report indicates no further related issues have been reported for this device.One year of service history was reviewed for this device with no problems identified related to the reported condition.Root cause: the root cause was the adjustment of the iv pole release mechanism.Correction: a trima field action has been initiated to notify all trima users to use precaution while transporting the device and a caution statement was included in the operator's manual.Corrective action: an internal capa has been initiated to evaluate reports of the iv pole dropping down suddenly.
 
Event Description
The customer reported that the iv pole on the trima machine would not stay in the upright position.No injury was reported for this incident and no patient was connected at the time the iv pole was sliding down, therefore no patient information is reasonably known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer Contact
steve kern
10810 w. collins ave
lakewood, CO 80215
3032392246
MDR Report Key7022339
MDR Text Key92773923
Report Number1722028-2017-00441
Device Sequence Number1
Product Code GKT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK010046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Notification
Type of Report Initial
Report Date 11/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number917000000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/19/2017
Initial Date FDA Received11/10/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/28/2003
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-