Model Number 7209808-
Device Problems Break (1069); Device Difficult to Setup or Prepare (1487)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/04/2017
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter: occurred during a procedure but before the use of the machine in the procedure.The hand piece is unable to plug in due to a broken port.There was no patient involvement, no patient injury, and no medical intervention required.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Complaint of broken hand piece receptacle was confirmed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.The root cause of the observed condition was determined to be a result of a misuse of the product.The instructions for use recommends "careful inspection of the operative control unit before and after the procedure for possible signs of damage.Immediate detection and repair of minor damage will extend the life of the operative control unit.¿ damage may occur if the control unit is handled roughly during use.Excessive forces may result in traumatic insertion, chipping particles of the optical glass, or other damage to the unit.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUCLEAR
Type of Device
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
MITG-OKLAHOMA CITY
75 s. meridian ave
oklahoma OK 73107
Manufacturer (Section G)
MITG-OKLAHOMA CITY
75 s. meridian ave
oklahoma OK 73107
Manufacturer Contact
sharon murphy
60 middletown ave
north haven, CT 06473
2034925267
MDR Report Key7024237
MDR Text Key92215716
Report Number1643264-2017-21130
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7209808-
Device Catalogue Number7209808-
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/16/2017
Initial Date FDA Received11/13/2017
Supplement Dates Manufacturer Received01/04/2018
Supplement Dates FDA Received01/22/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-